Calibration of ExoMars 2020 GTS (Ground Temperature Sensor) and its application on two band pyrometry

  • Miracle Israel Nazarious* (Corresponding Author)
  • , Maria-Paz Zorzano
  • , Javier Martin-Torres
  • *Corresponding author for this work
  • Luleå University of Technology
  • Centro de Astrobiologia (CSIC‐INTA)
  • Instituto Andaluz de Ciencias de la Tierra

Research output: Contribution to conferenceConference presentation (Unpublished paper)

15 Downloads (Pure)

Abstract

We describe the calibration procedure of the infrared thermopile used in the Ground Temperature Sensor (GTS) of the HABIT/ExoMars 2020 instrument, and report the preliminary performance test results at laboratory ambient and Martian equivalent environment. The setup and the procedures described in this work are easy to be implemented with any infrared thermopile sensors and facilitate developing a custom non- contact temperature/emissivity measurement system. We also developed a two-band pyrometer based on the calibration procedure to demonstrate the advantages of measuring the temperature and the emissivity of a remote target and detect the changes that are associated with the phases of the brine as hypothesized to be on the Martian surface. This remote detector can be of interest to the future surface exploration of Mars.
Original languageEnglish
Number of pages2
Publication statusPublished - 1 Sept 2019
EventEPSC-DPS Joint Meeting 2019 - Geneva, Switzerland
Duration: 15 Sept 201920 Sept 2019

Conference

ConferenceEPSC-DPS Joint Meeting 2019
Country/TerritorySwitzerland
CityGeneva
Period15/09/1920/09/19

Bibliographical note

EPSC Abstracts Vol. 13, EPSC-DPS2019-1330-2, 2019 EPSC-DPS Joint Meeting 2019

Fingerprint

Dive into the research topics of 'Calibration of ExoMars 2020 GTS (Ground Temperature Sensor) and its application on two band pyrometry'. Together they form a unique fingerprint.

Cite this