Skip to main navigation Skip to search Skip to main content

Correlation analysis of deep learning methods in S-ICD screening

  • Mohamed ElRefai* (Corresponding Author)
  • , Mohamed Abouelasaad
  • , Benedict M Wiles
  • , Anthony J Dunn
  • , Stefano Coniglio
  • , Alain B Zemkoho
  • , John Morgan
  • , Paul R Roberts
  • *Corresponding author for this work
  • University Hospital Southampton NHS Foundation Trust
  • University of Bergamo
  • Aberdeen Royal Infirmary

Research output: Contribution to journal › Article › peer-review

7 Downloads (Pure)

Abstract

BACKGROUND: Machine learning methods are used in the classification of various cardiovascular diseases through ECG data analysis. The concept of varying subcutaneous implantable cardiac defibrillator (S-ICD) eligibility, owing to the dynamicity of ECG signals, has been introduced before. There are practical limitations to acquiring longer durations of ECG signals for S-ICD screening. This study explored the potential use of deep learning methods in S-ICD screening.

METHODS: This was a retrospective study. A deep learning tool was used to provide descriptive analysis of the T:R ratios over 24 h recordings of S-ICD vectors. Spearman's rank correlation test was used to compare the results statistically to those of a "gold standard" S-ICD simulator.

RESULTS: A total of 14 patients (mean age: 63.7 ± 5.2 years, 71.4% male) were recruited and 28 vectors were analyzed. Mean T:R, standard deviation of T:R, and favorable ratio time (FVR)-a new concept introduced in this study-for all vectors combined were 0.21 ± 0.11, 0.08 ± 0.04, and 79 ± 30%, respectively. There were statistically significant strong correlations between the outcomes of our novel tool and the S-ICD simulator (p < .001).

CONCLUSION: Deep learning methods could provide a practical software solution to analyze data acquired for longer durations than current S-ICD screening practices. This could help select patients better suited for S-ICD therapy as well as guide vector selection in S-ICD eligible patients. Further work is needed before this could be translated into clinical practice.

Original languageEnglish
Article numbere13056
Number of pages11
JournalAnnals of noninvasive electrocardiology : the official journal of the International Society for Holter and Noninvasive Electrocardiology, Inc
Volume28
Issue number4
Early online date15 Mar 2023
DOIs
Publication statusPublished - Jul 2023

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • deep learning tools
  • screening
  • subcutaneous implantable cardiac defibrillator

Fingerprint

Dive into the research topics of 'Correlation analysis of deep learning methods in S-ICD screening'. Together they form a unique fingerprint.

Cite this