Abstract
The design, construction and use of a large-scale field-cycled proton- electron double-resonance imaging (FC-PEDRI) imager is described. The imager is based on a whole-body sized, vertical field, 59 mT permanent magnet. Field cycling is accomplished by the field compensation method, and uses a secondary, resistive magnet with an internal diameter of 52 cm. The magnetic field can be switched from zero to 59 mT or vice versa in 40 ms. It is used with a double-resonance coil assembly (NMR/EPR) comprising a solenoidal NMR transmit/receive coil and a coaxial, external birdcage resonator for EPR irradiation. Experiments to image the distribution of an exogenous nitroxide free radical in anaesthetized rabbits are described.
| Original language | English |
|---|---|
| Pages (from-to) | 1877-1886 |
| Number of pages | 10 |
| Journal | Physics in Medicine and Biology |
| Volume | 43 |
| Issue number | 7 |
| DOIs | |
| Publication status | Published - 30 Jul 1998 |
Bibliographical note
We thank Mr E Stevenson, Mr P Mullen and Mr J Carney for help in the construction of the screened room. This work was funded by the Medical Research Council and by the University of Aberdeen.Fingerprint
Dive into the research topics of 'Design, construction and use of a large-sample field-cycled PEDRI imager'. Together they form a unique fingerprint.Cite this