Abstract
Introduction
Haemodialysis patients who require defibrillator therapy are expected to benefit from the entirely avascular subcutaneous defibrillator (S-ICD), but haemodialysis is associated with dynamic changes in R and T wave amplitude which can impact S-ICD eligibility. A continuous assessment of S-ICD eligibility during haemodialysis has not previously been performed.
Material and methods
Continuous surface ECG recordings were obtained from a cohort of patients undergoing maintenance haemodialysis, but without an indication for an ICD. Automated vector screening was retrospectively performed at one-minute intervals throughout the dialysis session. Variations in S-ICD eligibility were calculated and in vectors with high degrees of variation, the underlying mechanism was identified.
Results
72 vector recordings (mean duration 254.1 ± 6.0 min) were obtained from 24 patients (mean age 64.3 ± 5.5 years, 68% male). At the start of haemodialysis 47 vectors were S-ICD eligible (65.2%). At the end of session, all of these vectors had remained eligible, and an additional 6 vectors had also become eligible (73.6%). High vector score variability was observed in 7 patients and the commonest cause was a progressive change in R:T ratio (71.5%).
Conclusion
In a haemodialysis population, a single haemodialysis session can be associated with a potential change in S-ICD eligibility in 8.4% of vectors, with up to 12.5% of vectors showing high degrees of variability, most commonly due to variations in R:T ratio. In an S-ICD population with similar characteristics S-ICD screening prior to haemodialysis would be expected to more accurately identify vectors that retain eligibility.
Haemodialysis patients who require defibrillator therapy are expected to benefit from the entirely avascular subcutaneous defibrillator (S-ICD), but haemodialysis is associated with dynamic changes in R and T wave amplitude which can impact S-ICD eligibility. A continuous assessment of S-ICD eligibility during haemodialysis has not previously been performed.
Material and methods
Continuous surface ECG recordings were obtained from a cohort of patients undergoing maintenance haemodialysis, but without an indication for an ICD. Automated vector screening was retrospectively performed at one-minute intervals throughout the dialysis session. Variations in S-ICD eligibility were calculated and in vectors with high degrees of variation, the underlying mechanism was identified.
Results
72 vector recordings (mean duration 254.1 ± 6.0 min) were obtained from 24 patients (mean age 64.3 ± 5.5 years, 68% male). At the start of haemodialysis 47 vectors were S-ICD eligible (65.2%). At the end of session, all of these vectors had remained eligible, and an additional 6 vectors had also become eligible (73.6%). High vector score variability was observed in 7 patients and the commonest cause was a progressive change in R:T ratio (71.5%).
Conclusion
In a haemodialysis population, a single haemodialysis session can be associated with a potential change in S-ICD eligibility in 8.4% of vectors, with up to 12.5% of vectors showing high degrees of variability, most commonly due to variations in R:T ratio. In an S-ICD population with similar characteristics S-ICD screening prior to haemodialysis would be expected to more accurately identify vectors that retain eligibility.
| Original language | English |
|---|---|
| Pages (from-to) | 21-27 |
| Number of pages | 6 |
| Journal | Journal of Electrocardiography |
| Volume | 72 |
| Early online date | 2 Mar 2022 |
| DOIs | |
| Publication status | Published - 1 May 2022 |
Bibliographical note
The authors would like to acknowledge the kindness and support that they received from all of the patients and staff at the Chandler's Ford Dialysis Unit.Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
Keywords
- S-ICD
- Subcutaneous
- Defibrillator
- Vector
- Haemodialysis
Fingerprint
Dive into the research topics of 'Electrocardiographic changes during haemodialysis and the potential impact on subcutaneous implantable cardioverter defibrillator eligibility'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS