Abstract
Object
This work aimed at evaluating the accuracy of MR-guided high-intensity focused ultrasound (MRgHIFU) brain therapy in human cadaver heads.
Methods
Eighteen heads of fresh human cadavers were removed with a dedicated protocol preventing intracerebral air penetration. The MR images allowed determination of the ultrasonic target: a part of the thalamic nucleus ventralis intermedius implicated in essential tremor. Osseous aberrations were corrected with simulation-based time reversal by using CT data from the heads. The ultrasonic session was performed with a 512-element phased-array transducer system operating at 1 MHz under stereotactic conditions with thermometric real-time MR monitoring performed using a 1.5-T imager.
Results
Dissection, imaging, targeting, and planning have validated the feasibility of this human cadaver model. The average temperature elevation measured by proton resonance frequency shift was 7.9°C ± 3°C. Based on MRI data, the accuracy of MRgHIFU is 0.4 ± 1 mm along the right/left axis, 0.7 ± 1.2 mm along the dorsal/ventral axis, and 0.5 ± 2.4 mm in the rostral/caudal axis.
Conclusions
Despite its limits (temperature, vascularization), the human cadaver model is effective for studying the accuracy of MRgHIFU brain therapy. With the 1-MHz system investigated here, there is millimetric accuracy.
This work aimed at evaluating the accuracy of MR-guided high-intensity focused ultrasound (MRgHIFU) brain therapy in human cadaver heads.
Methods
Eighteen heads of fresh human cadavers were removed with a dedicated protocol preventing intracerebral air penetration. The MR images allowed determination of the ultrasonic target: a part of the thalamic nucleus ventralis intermedius implicated in essential tremor. Osseous aberrations were corrected with simulation-based time reversal by using CT data from the heads. The ultrasonic session was performed with a 512-element phased-array transducer system operating at 1 MHz under stereotactic conditions with thermometric real-time MR monitoring performed using a 1.5-T imager.
Results
Dissection, imaging, targeting, and planning have validated the feasibility of this human cadaver model. The average temperature elevation measured by proton resonance frequency shift was 7.9°C ± 3°C. Based on MRI data, the accuracy of MRgHIFU is 0.4 ± 1 mm along the right/left axis, 0.7 ± 1.2 mm along the dorsal/ventral axis, and 0.5 ± 2.4 mm in the rostral/caudal axis.
Conclusions
Despite its limits (temperature, vascularization), the human cadaver model is effective for studying the accuracy of MRgHIFU brain therapy. With the 1-MHz system investigated here, there is millimetric accuracy.
| Original language | English |
|---|---|
| Pages (from-to) | 1046-1052 |
| Number of pages | 7 |
| Journal | Journal of Neurosurgery |
| Volume | 118 |
| Issue number | 5 |
| DOIs | |
| Publication status | Published - 1 May 2013 |
| Externally published | Yes |